Typical Parameters
Reference values only. The BMR, product specification and validated process always take precedence.
| Parameter | Typical value | Notes |
|---|---|---|
| Blender fill level | 40–60% of gross volume | About 50% is common. |
| Tumble blender speed | 10–30 rpm | Larger blenders run slower. |
| Main blend time | 10–20 minutes | As validated. |
| Lubrication blend time | 2–5 minutes | As validated. |
| Blend uniformity | Individual 90–110% of mean; RSD ≤ 5% | Or per specification. |
| Uniformity sample locations | ≥ 10 | Top, middle, bottom; legs and apex. |
| Uniformity sample size | 1–3 × unit dose weight | |
| Drum QA sample | ≈ 50 g per drum | Or per sampling plan. |
| Moisture sample size | 2–5 g | Per method. |
| Density cylinder | 100 or 250 mL | |
| Tapped density taps | 10 / 500 / 1,250 (+1,250) | 3 mm drop, 250 taps/min. |
| Good flow | Carr ≤ 15%; Hausner ≤ 1.18 | |
| Balance warm-up | 30–60 minutes | |
| Balance room | 20 ± 2 °C, 45–55% RH | |
| Calibration weights | OIML E2 or F1 | Accredited traceability. |
| Small vibratory screen load | < 3 kg | Per manufacturer. |