Guide9.4
Analytical balance calibration
| Step | Action | Why |
|---|---|---|
| 1. Prepare | Place on a stable, vibration-free bench; level with the bubble indicator; connect power; warm up 30–60 minutes. | Electronics need thermal and electrical stability or results drift. |
| 2. Clean | Brush the pan, draft shield and chamber with a soft brush. | Debris adds weight and distorts the calibration. |
| 3. Enter calibration mode | With the pan empty and doors closed, press CAL or select Menu > Calibration > Internal/External. | Objects or air currents would corrupt the baseline. |
| 4. Zero | Wait for a stable zero, then confirm. | Establishes the true zero reference. |
| 5. Apply span weight | Place the certified weight (e.g. 100 g) in the centre of the pan with tongs; close the doors. | Off-centre loading changes the reading; skin oils and heat change the weight’s mass. |
| 6. Confirm | When the stable reading matches the weight (e.g. 100.0000 g), accept. | Locks in the span value. |
| 7. Remove weight | Remove with tongs, close doors, confirm return to zero. | Confirms the balance resets correctly. |
| 8. Save | Select Save/Store; the balance returns to weighing mode. | Stores the new calibration internally. |
| 9. Record | Print the report and record results in the calibration log. | Documented evidence for GMP and audit. |
| 10. Label | Attach a calibration label with the next due date. | Shows the balance is fit for use. |